GENERAL
Monitor Performance Specifications |
| Display, Resolution |
8.4" color TFT, 800 x 600 pixels |
| Dimension, Weight |
238 (w) x 250 (h) x 163 (d) mm.
Approx. 3.1 kg |
| Indicators |
Up to 3 waves (ECG, SpO2, Respiration)
Categorized alarms (3 priority levels)
Visual alarm lamp
Heart rate, SpO2 tone
Battery status
External power LED
|
| Interfaces |
DC input connector: 18 VDC, 2.5A
Defib Sync Output
- Signal Level: 0 to 5 V pulse
- Pulse Width: 100 +/- 10ms
LAN digital output for transferring data
VGA output
|
| Battery |
Bettery Type: Li-ion
Battery status indicator
Operating Time: 2 hours typically (fully charged) |
| Thermal Printer (optional) |
Speeds: 25, 50 mm/sec
Paper width: 58 mm |
| Graphical and Tabular Trends |
| Tabular Trends |
Memory Storage: 24 hours
Data Interval: 1, 5, 15, 30 min., and 1 hour
Tabular Format: One table for all variables |
| Graphical Trends |
Display Duration: 30, 60, 90 min, 3, 6, 12 hours |
| Accessories |
| Standard Accessories |
3- lead ECG cable - 1 ea.
Electrodes - 10 ea.
NIBP extension tube, 3m long - 1 ea.
Adult Cuff, 25-35 cm, reusable - 1 ea.
SpO2 sensor extension cable (2m) - 1 ea.
SpO2 sensor reusable - 1 ea.
DC adapter, 18VDC, 2.5A - 1 set |
| Optional Accessories |
Temperature probe, Surface / Skin, reusable
5 lead patient cable
Neonate cuff
Child cuff
|
| Data Storage |
120 Hours trends
20 cases of 10 sec alarm waveform |
| Language |
English, French, Spanish, Italian, German, Chinese |
*Interpretive results are intended only as guidance for the qualified physicians and must not be relied upoon as diagnosis |
PERFORMANCE
ECG |
| Leads |
3 lead, 5 leads(optional) |
| Heart Rate Range |
30 - 300 bpm |
| Hearth Rate Accuracy |
+/- 3 bpm |
| Bandwidth |
0.5 Hz to 40 Hz |
| ECG Size (Sensitivity) |
0.5, 1, 2, 4 mV / cm |
| Pacemaker Detection Mode |
Indicator on waveform display, user selectable |
| Differential Input Impedance |
>5 mo |
| CMRR |
>90 db at 50 or 60 Hz |
| Input Dynamic Range |
+/- 5 mVAC, +/- 300 mVDC |
| Defribrillator Discharge |
<4 sec |
| Lead-Off Detection |
Yes |
| SpO2 |
|
| % Saturation Range |
0% to 100% |
| Saturation Accuracy |
70 to 100% +/- 2 digits
0 to 69% unspecified |
| Pulse Rate Range |
0 to 254 bpm |
| Pulse Rate Accuracy |
+/- 2 bpm |
| NIBP |
| Measurement Modes |
Manual: Single measurment
Auto: Automatic intervals of 1, 2, 3, 4, 5, 10, 15, 20, 30 min. and 1, 2, 4, 8, hr.
STAT: Continous measurement of 5 min |
| Cuff Presure Display |
0 to 300 mmHg |
| B.P. measurement range |
Adult
- Systolic: 50 to 260 mmHg
- Mean : 30 to 220 mmHg
- Diastolic: 20 to 200 mmHg
Pediatric
- Systolic: 20 to 230 mmHg
- Mean: 30 to 190 mmHg
- Diastolic: 20 to 170 mmHg
Neonate
- Systolic: 50 to 120 mmHg
- Mean: 30 to 100 mmHg
- Diastolic: 20 to 90 mmHg
|
| Respiration |
| Range |
5 to 120 breaths / min |
| Accuracy |
+/- 3 breaths/min |
| Temperature Unit |
| Range |
15° C to 45° C (59° F to 113° F) |
| Accuracy |
+/- 0. 1°C |
| Compatible with |
YS1 Series 400 temperature probes |