SPECIFICATIONS

GENERAL
Monitor Performance Specifications

Display, Resolution

8.4" color TFT, 800 x 600 pixels

Dimension, Weight

238 (w) x 250 (h) x 163 (d) mm. Approx. 3.1 kg

Indicators

Up to 3 waves (ECG, SpO2, Respiration)
Categorized alarms (3 priority levels)
Visual alarm lamp
Heart rate, SpO2 tone
Battery status
External power LED

Interfaces

DC input connector: 18 VDC, 2.5A
Defib Sync Output
- Signal Level: 0 to 5 V pulse
- Pulse Width: 100 +/- 10ms
LAN digital output for transferring data
VGA output

Battery

Bettery Type: Li-ion
Battery status indicator
Operating Time: 2 hours typically (fully charged)

Thermal Printer (optional)

Speeds: 25, 50 mm/sec
Paper width: 58 mm

Graphical and Tabular Trends
Tabular Trends Memory Storage: 24 hours
Data Interval: 1, 5, 15, 30 min., and 1 hour
Tabular Format: One table for all variables
Graphical Trends Display Duration: 30, 60, 90 min, 3, 6, 12 hours
Accessories
Standard Accessories

3- lead ECG cable - 1 ea.
Electrodes - 10 ea.
NIBP extension tube, 3m long - 1 ea.
Adult Cuff, 25-35 cm, reusable - 1 ea.
SpO2 sensor extension cable (2m) - 1 ea.
SpO2 sensor reusable - 1 ea.
DC adapter, 18VDC, 2.5A - 1 set

Optional Accessories

Temperature probe, Surface / Skin, reusable
5 lead patient cable
Neonate cuff
Child cuff

Data Storage 120 Hours trends
20 cases of 10 sec alarm waveform
Language

English, French, Spanish, Italian, German, Chinese

*Interpretive results are intended only as guidance for the qualified physicians and must not be relied upoon as diagnosis

PERFORMANCE
ECG
Leads

3 lead, 5 leads(optional)

Heart Rate Range

30 - 300 bpm

Hearth Rate Accuracy +/- 3 bpm
Bandwidth 0.5 Hz to 40 Hz
ECG Size (Sensitivity) 0.5, 1, 2, 4 mV / cm
Pacemaker Detection Mode

Indicator on waveform display, user selectable

Differential Input Impedance

>5 mo

CMRR

>90 db at 50 or 60 Hz

Input Dynamic Range +/- 5 mVAC, +/- 300 mVDC
Defribrillator Discharge

<4 sec

Lead-Off Detection Yes
SpO2  
% Saturation Range 0% to 100%
Saturation Accuracy 70 to 100% +/- 2 digits
0 to 69% unspecified
Pulse Rate Range 0 to 254 bpm
Pulse Rate Accuracy +/- 2 bpm
NIBP
Measurement Modes Manual: Single measurment
Auto: Automatic intervals of 1, 2, 3, 4, 5, 10, 15, 20, 30 min. and 1, 2, 4, 8, hr.
STAT: Continous measurement of 5 min
Cuff Presure Display

0 to 300 mmHg

B.P. measurement range

Adult
- Systolic: 50 to 260 mmHg
- Mean : 30 to 220 mmHg
- Diastolic: 20 to 200 mmHg
Pediatric
- Systolic: 20 to 230 mmHg
- Mean: 30 to 190 mmHg
- Diastolic: 20 to 170 mmHg
Neonate
- Systolic: 50 to 120 mmHg
- Mean: 30 to 100 mmHg
- Diastolic: 20 to 90 mmHg

Respiration
Range 5 to 120 breaths / min
Accuracy +/- 3 breaths/min
Temperature Unit
Range 15° C to 45° C (59° F to 113° F)
Accuracy +/- 0. 1°C
Compatible with YS1 Series 400 temperature probes