GENERAL
Monitor Performance Specifications |
| Display, Resolution |
7" Wide color TFT, 800 x 480 pixels |
| Dimension, Weight |
238 (w) x 250 (h) x 163 (d) mm.
Approx. 3.1 kg |
| Trace |
4 waveforms: 2*ECG, Sp02, Respiration
Sweep speed: 6.25, 12.5, 25, 50 mm/sec |
| Indicators |
Categorized alarms (3 priority levels)
Heart beat tone
Spo2 pulse pitch tone
Battery status
External power LED |
| Interfaces |
DC input connector: 12 to 16 VDC, 2.5A
Defib Sync Output
- Signal Level: 0 to 5 V pulse
- Pulse Width: 100 +/- 10ms
LAN digital output for transferring data
Nurse call system connection
- 0.3A at 125VAC
- 1A at 24VAC
DC output: 5VDC, 1A Max
|
| Battery |
Rechargeable Li-ion battery
2 hours of continuous operation |
| Thermal Printer (option) |
Speeds: 25, 50 mm/sec,
Paper width: 58 mm |
| Data Storage |
120 Hours trends
20 cases of 10 sec alarm waveform |
| Language |
English, French, Spanish, Italian, German, Chinese, Russian, Czech, Polish, Turkish, Romanian, Portuguese, Bulgarian |
| Accessories |
| Standard Accessories |
3- lead ECG cable - 1 ea.
ECG extension cable - 1 ea.
NIBP extension tube, 3m long - 1 ea.
NIBP vet cuff neonate disposable #1-#5 - 1 set
NIBP vet cuff infant reusable - 1 ea.
SpO2 extension cable (2m) - 1 ea.
Reusable multisite SpO2 probe - 1 ea.
B-Link automated record keeping software
DC adapter, 18VDC, 2.5A - 1 set |
| Optional Accessories |
5 lead vet ECG cable
Temperature probe (Skin)
Temperature probe (Rectal) |
PERFORMANCE
ECG |
| Lead Type |
3 lead, 5 lead |
| Lead Selection |
3-lead : I, II, III
5-lead : I, II, III, aVR, aVL, aVF, V |
| ECG waveforms |
3-lead : 1 channel : 2 traces
5-lead : 2 channels |
| Heart Rate Range |
30 - 300 bpm |
| Hearth Rate Accuracy |
+/- 1bpm or +/- 1%, whichever is greater |
| Sweep Speed |
6.25, 12.5, 25, 50 mm/sec |
| Filter |
Dianosis : 0.1 ~ 150HZ
Monitoring : 0.1 ~ 40Hz
Moderate : 0.1 ~
20Hz
Maximum : 5 ~ 20Hz |
| S-T segment detection range |
-2.0 to 2.0 mV |
| Arrhythmia Analysis |
ASYSTOLE, VTACH, VFIB |
| Pacemaker Detection Mode |
Indicator on waveform display, user selectable |
| Protection |
Against selectrosurgical interference and defibrillation |
| Respiration |
| Method |
Thoracic impedance |
| Measurement Range |
5 - 120 Breath per minute |
| Accuracy |
+/- 1 breaths/min |
| Apnea Alarm |
Yes |
| SpO2 |
|
| Saturation Range |
0% to 100% |
| Saturation Accuracy |
70 to 100% +/- 2 digits
0 to 69% unspecified |
| Pulse Rate Range |
0 to 254 bpm |
| Pulse Rate Accuracy |
+/- 2 bpm |
| Temperature |
| Measurement Range |
15° C to 45° C (59° F to 113° F) |
| Accuracy |
+/- 1°C |
| Compatible with |
YS1 Series 400 temperature probes |
| NIBP |
| Method |
Automatic Oscillometric |
| Operation Mode |
Manual / Automaitc / Continuous |
| Cuff Pressure Display |
30 to 300mmHG |
| Measurement range |
Preasure range : 20 to 260mmHg |
| Accuracy |
mean error : less than +/- 5 mmHg
standard deviation : less than 8 mmHg |