GENERAL
Monitor Performance Specifications

Display

10.4" color TFT screen (800 x 600)

Dimension

270 (w) x 310 (h) x 172 (d) mm.
Approx. 4 kg (8.8 Lbs.)

Trace

5 waveforms (ECG, SpO2, 2 IBP, RR or EtCO2)

Sweep Speed

6.25, 12.5, 25, 50 mm/sec

Indicators

Categorized alarms (3 priority levels)
Visual alarm lamp handle
QRS Beep and alarm sound
Battery status
External power LED

Interfaces

LAN digital output for transferring data
VGA Out
Defib Sync Output
- Signal Level: 0 to 5 V pulse
- Pulse Width: 100 +/- 10ms
AC input : 100 ~ 240V, 50/60 Hz, 1.2A
DC input : 12 to 18Vdc, 2.5A Max.
DC output : 3.3Vdc, 700mA Max.

Battery

Internal battery: Li-ion
Battery status indicator
Operating Time: 2.5 hours typically (fully charged battery)

Thermal Printer
(Optional)

Built-in, 3 channels
Speeds: 25, 50 mm/sec
Paper width: 58 mm

Graphical and Tabular Trends
Tabular Trends

Memory Storage: 24 hours
Data Interval: 1, 5, 15, 30 min., and 1 hour
Tabular Format: One table for all variables

Graphical Trends

Display duration: 30, 60, 90 min., 3, 6, 12 hours

Accessory
Standard Accessory

3- lead ECG cable and clips - 1 ea.
Electrodes - 10 ea.
NIBP extension tube - 1 ea.
NIBP Neonate Cuff #1~#5 - 1 ea.
SpO2 sensor extension cable - 1 ea.
SpO2 sensor - 1 ea.
EtCO2 FilterLine set - 1 ea.
Power Transformer (DC Adaptor) - 1 ea.
pwoer Cable - 1 ea.

Optional Accessory

Temperature probe Surface/Skin, reusable
Thermal paper
5-lead patient cable
IBP disposable preasure transducer kit
IBP logic pressure transducer kit

PERFORMANCE
ECG
Leads

3 leads or 5 leads selectable

Heart Rate Range

30 to 300 bpm

Hearth Rate Accuracy +/- 3 bpm
Bandwidth

0.5 Hz to 40 Hz

Sweep Speeds

6.25, 12.5, 25, 50 mm/sec

ECG Size Sensitivity

0.5, 1, 2, 4 mV/cm

Pacemaker Detection Mode

Indicator on waveform display, user selectable

Differential Input Impedance >5MQ
CMRR

>90dB at 50 or 60 Hz

Input Dynamic Range

+/- 5mVAC, +/- 300 mVDC

Defibrillator Discharge

< 4 sec

Arrhythmia analysis VTAC/VFIB, VTAC, ASYSTOLIC
Lead-Off Detection with display indicator
SpO2  
% Saturation Range 0% to 100%
Pulse Rate 30 to 254 bpm
SpO2 Accuracy - 70% to 100% +/- 2 digits
- 0% to 69% unspecified
NIBP
Method Oscillometric
Measurement Modes

Manual, Automatic, STAT

Cuff Pressure Display 0 to 300mmHg in 1mmHg step
B.P. measurement range Systolic: 50 to 270 mmHg
Mean Arterial Pressure: 40 to 245 mmHg
Diastolic: 30 to 220 mmHg
Over pressure protection Yes
Adjustable Cuff Inflation Pressure
Respirations
Range 5 to 120 breaths/min
Accuracy +/- 3 breaths/min
Sweep Speeds 6.25, 12.5, 25, 50 mm/sec
Apnea alarm 0 to 30 sec
Temperature Unit
Range 15° C to 45° C (59° F to 113° F)
Accuracy - 25°C to 45° C +/- 0.1°C
- 15°C to 24°C +/- 0.2°C
Compatible with YS1 Series 400 temperature probes
IBP
Channels 2
Measuring Range -50 to 300 mmHg
Measuring Accuracy +/- 1 mmHg +/- 1 digit (<100 mmHg)
+/- 1 % +/- 1 digit (>100 mmHg)
Pulse rate counting range 0 to 300 beats/min
Zero balancing range +/- 200 mmHg
Accuracy +/- 1 mmHg
Drift +/- 1 mmHg over 24 hours
Transducer sensitivity 5_v/V/mmHg
EtCO2
Method Infrared Absorption
Measurement Mode MicroStream
Measurement Range 0 to 99 mmHg
Respiration Range 0 to 150 bpm
Accuracy +/- 2 mmHg, 0 to 40 mmHg
+/- 5 mmHg, 41 to 76 mmHg
+/- 10 mmHg, 77 to 99 mmHg
Resolution 1mmHg
Apnea alarm Yes