Spirometer Sensor
SPM-300
SPM-300 is a spirometer sensor which works with the Cardio7-S and SpiroCare. SPM-300 uses a disposable sensor, so it is very easy to maintain because the sensor does not require disinfecting between uses. These disposable sensors make an accurate measurement possible every time.
FEATURES
- FVC, SVC, and MVV Test
- Pre- and Post-Bronchodilator Comparison
- Using a Disposable Sensor
- Working with the Cardio7-S and SpiroCare
It is highly recommended to calibrate a spirometer with Bionet 3L calibration syringe periodically, in order to ensure accurate calibration of your spirometer.
Our calibration recommendations are based upon guidelines from the American Thoracic Society (ATS) and the European Respiratory Society (ERS)
- The volume accuracy of the spirometer must be checked at least daily with a single discharge of a 3-L calibrated syringe.
- At least quarterly, volume spirometers must have their calibration checked over their entire volume range using a calibrated syringe or an equivalent volume standard. The measured volume should be within 3.5% of the reading or 65 mL, whichever is greater. This limit includes the 0.5% accuracy limit for a 3-L syringe.
SPM-300 is a spirometer sensor which works with the Cardio7-S and SpiroCare. SPM-300 uses a disposable sensor, so it is very easy to maintain because the sensor does not require disinfecting between uses. These disposable sensors make an accurate measurement possible every time.
SPM-300 is a spirometer sensor which works with the Cardio7-S and SpiroCare.
By adding the SPM-300 to the Cardio7, you can test FVC, SVC, MVV and perform a pre-post bronchodilator comparison. Real-time volume-flow and time-flow monitoring is provided on the color LCD display. 120 tests are stored and a full size printout is provided for easy charting.
The disposable mouthpiece, SmarTube, is used. This makes it very easy to maintain the system because the sensor does not require disinfecting between uses. These disposable sensors make an accurate measurement possible every time.
wdt_ID | Feature | Specification |
---|---|---|
47 | Dimension | 47 (W) x 200 (H) x 34 (D)mm, approx. 250g |
48 | Measuring Values | - FVC: FVC, FEV 1.0, FEV 1/FVC, FEF 0.2 - 1.2L, FEF 25 - 75%, FEF 75 - 85%, PEF, FEF 25%, FEF 50%, FEF 75%, FIVC, FEV6, PEFT, FET 100% - SVC: SVC, ERV, IRV, TV, EC - MVV: MVV, FB, TV |
56 | Presentation | Flow volume loop MVV: MVV, FB, TV Measurement values table |
59 | Measuring Range | Flow: 0 to ± 14 L/s Volume: 0 to ± 12 L |
62 | Measuring Method | Differential pressure method |
65 | Prediction Equation | Morris-Polgar, ECCS-Quanjer, Knudson-ITS, Pereira, Korea CJK |
66 | Sample Rate | 200 samples/sec |
67 | Flow Impedance | < 0.2 mbar s/l at 12 l/s |
68 | Measuring Accuracy | Complies with ATS (American Thoracic Society) |
69 | Environment | Ambient temperature: 15 to 40°C (59 to 104°F) Relative humidity: 10 to 90% (non-condensing) Atmospheric pressure: 700 to 1060 hPa |
* Specifications can be changed without prior notification.
Spirometer FVC Test (SPM-300 + Cardio7)
SPM-300(Bionet Spirometer) + Cardio7 Calibration
- Overview
-
It is highly recommended to calibrate a spirometer with Bionet 3L calibration syringe periodically, in order to ensure accurate calibration of your spirometer.
Our calibration recommendations are based upon guidelines from the American Thoracic Society (ATS) and the European Respiratory Society (ERS)
- The volume accuracy of the spirometer must be checked at least daily with a single discharge of a 3-L calibrated syringe.
- At least quarterly, volume spirometers must have their calibration checked over their entire volume range using a calibrated syringe or an equivalent volume standard. The measured volume should be within 3.5% of the reading or 65 mL, whichever is greater. This limit includes the 0.5% accuracy limit for a 3-L syringe.
SPM-300 is a spirometer sensor which works with the Cardio7-S and SpiroCare. SPM-300 uses a disposable sensor, so it is very easy to maintain because the sensor does not require disinfecting between uses. These disposable sensors make an accurate measurement possible every time.
SPM-300 is a spirometer sensor which works with the Cardio7-S and SpiroCare.
By adding the SPM-300 to the Cardio7, you can test FVC, SVC, MVV and perform a pre-post bronchodilator comparison. Real-time volume-flow and time-flow monitoring is provided on the color LCD display. 120 tests are stored and a full size printout is provided for easy charting.
The disposable mouthpiece, SmarTube, is used. This makes it very easy to maintain the system because the sensor does not require disinfecting between uses. These disposable sensors make an accurate measurement possible every time.
- Specifications
-
wdt_ID Feature Specification 47 Dimension 47 (W) x 200 (H) x 34 (D)mm, approx. 250g 48 Measuring Values - FVC: FVC, FEV 1.0, FEV 1/FVC, FEF 0.2 - 1.2L, FEF 25 - 75%, FEF 75 - 85%, PEF, FEF 25%, FEF 50%, FEF 75%, FIVC, FEV6, PEFT, FET 100%
- SVC: SVC, ERV, IRV, TV, EC
- MVV: MVV, FB, TV56 Presentation Flow volume loop
MVV: MVV, FB, TV
Measurement values table59 Measuring Range Flow: 0 to ± 14 L/s
Volume: 0 to ± 12 L62 Measuring Method Differential pressure method 65 Prediction Equation Morris-Polgar, ECCS-Quanjer, Knudson-ITS, Pereira, Korea CJK 66 Sample Rate 200 samples/sec 67 Flow Impedance < 0.2 mbar s/l at 12 l/s 68 Measuring Accuracy Complies with ATS (American Thoracic Society) 69 Environment Ambient temperature: 15 to 40°C (59 to 104°F)
Relative humidity: 10 to 90% (non-condensing)
Atmospheric pressure: 700 to 1060 hPa* Specifications can be changed without prior notification.